01
How to do a serial dilution
- Label the tubes and add the diluent volume to each.
- Transfer the calculated volume from the stock into tube 1 and mix thoroughly.
- With a fresh tip, transfer the same volume from tube 1 into tube 2 and mix. Repeat down the row.
- The last tube holds extra volume (nothing is taken out of it); discard the transfer volume from it if every tube must be equal.
02
Serial dilution formulas
transfer volume = tube volume ÷ dilution factor
diluent volume = tube volume − transfer volume
concentration in tube n = stock ÷ factorⁿ
If you know the range you need instead of the factor, choose “Lowest concentration I need in the last tube”: the factor is then (stock ÷ lowest)^(1 ÷ number of tubes). Going from 1,000 to 1 in six tubes needs a factor of 3.16 per step.
“Tube volume” is the volume in a tube just after the transfer in. The dilution factor is total volume divided by the volume transferred: a 10-fold step is 1 part sample in 10 parts total.
03
Worked example: a tenfold series
The calculator opens with a 1 mg/mL stock, factor 10, six tubes of 1,000 µL: 900 µL of diluent in each tube and 100 µL transferred.
| Tube | Dilution | Concentration |
|---|---|---|
| 1 | 1:10 | 0.1 mg/mL |
| 2 | 1:100 | 0.01 mg/mL |
| 3 | 1:1,000 | 0.001 mg/mL |
| 4 | 1:10,000 | 1 × 10⁻⁴ mg/mL |
| 5 | 1:100,000 | 1 × 10⁻⁵ mg/mL |
| 6 | 1:1,000,000 | 1 × 10⁻⁶ mg/mL |
04
Common series
| Series | Factor | For 200 µL per tube | Typical use |
|---|---|---|---|
| Doubling (2-fold) | 2 | 100 µL into 100 µL | MIC assays, antibody titres |
| 3-fold | 3 | 66.7 µL into 133.3 µL | Dose–response curves |
| Tenfold | 10 | 20 µL into 180 µL | Plate counts, qPCR standards |
05
Keeping the error small
- Errors multiply down a series: a 3% pipetting error per step becomes about 20% after six steps. Mix each tube well and change tips every time.
- Avoid transfer volumes below what your pipette handles accurately; raise the tube volume instead.
- For a standard curve, fit the measured values with the linear regression calculator.
06
Common serial dilution mistakes
- Adding the transfer to the full volume of diluent. 100 µL into 1,000 µL is an 11-fold step; after six tubes the dilution is 1,771,561, not 1,000,000, and the last tube is 44% weaker than labelled.
- Letting small errors run. A 5% error at each step, all in the same direction, grows to 63% after ten tubes, because the errors multiply.
- Forgetting the last tube. Nothing is taken out of it, so it holds 1,000 µL while the others hold 900 µL; remove 100 µL if every tube must match.
Trust
How this page is checked
- The calculation runs on code with automated tests against worked reference values. Every number in the examples on this page was produced by that same code, not typed by hand.
- The formulas and conventions used are stated on the page, so you can reproduce the result yourself or compare it with your own software.
- Last reviewed by the Sciloom team, Zebra Nine Studio Inc. It is a tool for research and teaching, not a substitute for a statistician, a validated lab method or medical advice.
Cite this page
Sciloom. (2026). Serial dilution calculator. Zebra Nine Studio Inc. https://sciloom.si/serial-dilution-calculator
Frequently asked questions
How do I calculate a serial dilution?
Divide the volume you want in each tube by the dilution factor to get the transfer volume; the rest is diluent. Each tube’s concentration is the previous one divided by the factor.
What is a 1:10 serial dilution?
A series in which each tube is ten times more dilute than the one before: one part from the previous tube plus nine parts diluent, repeated.
How do I find the final dilution factor?
Multiply the factors of every step. Three tenfold steps give 10 × 10 × 10 = 1,000, written 1:1,000 or 10⁻³.
Why change the pipette tip at each step?
Liquid clinging to the tip carries extra sample forward, which makes later tubes more concentrated than calculated.