PCR efficiency calculator

This PCR efficiency calculator converts the slope of a qPCR standard curve into amplification efficiency. Paste the quantities of a dilution series with their Ct values and it fits the line for you, or enter a slope you already have.

Your numbers are calculated in your browser. Nothing you enter is sent or stored.

PCR efficiency calculatorOr describe it in Studio →

Quantities can be copies, ng or relative amounts (1, 0.1, 0.01). 1e5 means 100,000.

Amplification efficiency

98%

Within the usual 90 to 110% acceptance range.

Amplification factor
1.98 per cycle
Slope
-3.37
Intercept
35.32
R²
0.99997
Points
5

To report

E = 10^(−1/-3.37) − 1 = 98%, R² = 0.99997

Check your result: each tenfold step in the dilution series should add about 3.3 cycles. In the example the steps are 3.4, 3.3, 3.4 and 3.4, the slope is −3.37 and the efficiency 98.0%. The amplification factor shown should equal 1 + efficiency ÷ 100: 1.980.
range the result is checked against
90–110%
range the result is checked against
R² below which the tool warns
0.98
R² below which the tool warns
efficiency of the example dilution series
98.0%
efficiency of the example dilution series
points needed for a fit
3
points needed for a fit
  • Efficiency from slope
  • Fits a dilution series
  • Slope, intercept and R²
  • Checks the 90–110% range
  • Runs in your browser, free

01

PCR efficiency formula

E = 10^(−1 ÷ slope) − 1

efficiency (%) = E × 100

The slope comes from plotting Ct against log₁₀ of the starting quantity. A perfect reaction doubles the product every cycle, which gives a slope of −3.32 and an efficiency of 100%.

02

Slope to efficiency

SlopeEfficiencyVerdict
−3.10110.2%just above range
−3.32100.0%ideal
−3.3798.0%in range
−3.6089.6%just below range

The commonly accepted window is 90 to 110%, roughly slopes from −3.6 to −3.1, together with an R² of at least 0.98 for the standard curve.

03

Worked example

A tenfold dilution series from 10⁶ to 10² copies gives mean Ct values of 15.1, 18.5, 21.8, 25.2 and 28.6.

fitted line: Ct = −3.37 × log₁₀(quantity) + 35.32, R² = 0.99997

E = 10^(1 ÷ 3.37) − 1 = 0.980 → 98.0%

04

Common mistakes when measuring PCR efficiency

  • Converting a natural-log slope with the log₁₀ formula. The example slope of −3.37 per log₁₀ unit is −1.46 per natural-log unit, and that slope put into E = 10^(−1 ÷ slope) − 1 gives 382% instead of 98%.
  • Typing the logarithms as quantities. Entering 6, 5, 4, 3, 2 instead of 10⁶ to 10² takes the log a second time and the example data come out at 8.6%.
  • Labelling a twofold series as tenfold. Ct values 15, 16, 17, 18 and 19 from a twofold series, entered against tenfold quantities, give a slope of −1 and an efficiency of 900%.
  • Trusting R² alone. A perfectly straight series with 3.0 cycles per tenfold step has R² = 1 and an efficiency of 115.4%, outside the 90 to 110% range.

Trust

How this page is checked

  • The calculation runs on code with automated tests against worked reference values. Every number in the examples on this page was produced by that same code, not typed by hand.
  • The formulas and conventions used are stated on the page, so you can reproduce the result yourself or compare it with your own software.
  • Sources: Pfaffl (2001), A new mathematical model for relative quantification in real-time RT-PCR.
  • Last reviewed by the Sciloom team, Zebra Nine Studio Inc. It is a tool for research and teaching, not a substitute for a statistician, a validated lab method or medical advice.

Cite this page

Sciloom. (2026). PCR efficiency calculator. Zebra Nine Studio Inc. https://sciloom.si/pcr-efficiency-calculator

Read the full methodology · Who makes Sciloom

Frequently asked questions

What is a good qPCR efficiency?

Between 90% and 110%. Values in that range mean the product roughly doubles each cycle and the 2^−ΔΔCt method can be used.

How can efficiency be above 100%?

It cannot be in reality. An apparent value above 100% usually comes from inhibitors in the most concentrated standards, primer dimers, or pipetting errors along the dilution series.

How many dilution points do I need?

At least five points across several orders of magnitude, each run in triplicate, is the usual recommendation. The calculator needs three or more.

Do I enter copies or dilutions?

Either. Only the ratio between points matters for the slope, so relative amounts such as 1, 0.1 and 0.01 give the same efficiency as copy numbers.

Efficiency in range? Go on to the delta delta Ct calculator for fold change.
Free to start, no sign-up.
Back to the tool ↑

More research tools